405C Performance Specification |
|
Readout Ranges
Sodium 0 - 180 mmol/l (in 2 mmol/l increments)
Potassium 0 - 10.0 mmol/l (in 0.1 mmol/l increments)
Lithium 0 - 3.0 mmol/l (in 0.1 mmol/l increments)
|
Display Resolution
| Sodium: | 2 | mmol/l |
| Potassium: | 0.1 | mmol/l |
| Lithium: | 0.1 | mmol/l |
Linearity
| Sodium: | ±2 | mmol/l |
| Potassium: | ±0.2 | mmol/l |
| Lithium: | ±0.1 | mmol/l |
Reproducibility
Based upon the measurement of 20 replicates over a 10 minute period, reproducibility will be better than:
| Sodium: | (at 140 mmol/l level) | 1.5% C.V. |
| Potassium: | (at 5.0 mmol/l level) | 1.5% C.V. |
| Potassium: | (at 80 mmol/l level) | 1.5% C.V. |
| Lithium: | (at 2.5 mmol/l level) | 2.5% C.V. |
Recorder Output
DC 100mV
|
Sample Requirement
| Mode | Test | Sample Vol |
| Serum | Na | 10 (Micro Litre) |
| Urine | Na | 10 (Micro Litre) |
| Serum | K | 50 (Micro Litre) |
| Urine | K | 10 (Micro Litre) |
| Serum | Li | 100 (Micro Litre) |
Diluted sample aspiration rate
Aspiration rate: 2 - 6 ml/min
Time to steady state reading
(from presentation of sample):
typically 8 seconds.
Applications |
Flame photometry is well proven in clinical
applications and is universally accepted as
the reference method.
The 405C will economically fulfill the
requirements of:
Hospital Clinical Chemistry laboratories
Renal Dialysis laboratories
Independent Clinical Testing laboratories
Veterinary laboratories
Universities, Medical schools and
technical colleges
|
|