Operating Sequence |
A typical calibration and operating sequence
is shown below:
- Switch on power and fuel.
(compressor starts automatically.)
- Press ignition switch.
- Select filter.
- Dilute standards and samples.
- Aspirate a blank and set zero.
- Aspirate standard solution and
set sensitivity for correct reading.
- Aspirate unknown and record results.
|
General Description |
|
Controls and indicators
Power on/off
Power indicator
Display (panel meter 112x56mm)
Fuel (rotary control)
Blank (rotary control)
Sensitivity coarse (rotary control)
Air regulator(fitted at rear)
|
|
Recorder output
DC 100mv
Air Supply
Oil free 6 litres/min at 14 psi minimum
(Fater Electronic model 850 compressors
are recommended.)
Fuel Supply
High grade propane,butane or propane/butane
mixture. Fuel should be free from heavy
hydrocarbon deposits and regulated to
approx, 30 psi at the cylinder.
Natural gas at 3" - 10" W.G.
Electrical Supply
198 to 264 volts AC
50/60 Hz.
Size and Weight
550 mm(height) 420mm(width)
282mm(depth)- 11.8kg
Packed 730mm(height) 552mm(width)
450mm(depth)- 16.5kg
|